Absorption Measurements in Liquid Core Waveguides Using Cavity
Loading...
Authors
Bescherer, Klaus
Barnes, Jack A.
Loock, Hans-Peter
Date
2016-03-07
Type
journal article
Language
en
Keyword
Absorption Measurements , Liquid Core Waveguides , Cavity Ring-Down
Alternative Title
Abstract
Short liquid core waveguides (LCWs) were included into a fiber-loop cavity ring-down absorption spectrometer to reduce the detection limit over, both, single pass absorption in a LCW and cavityenhanced absorption using a conventional fiber-loop cavity. LCWs of 5 and 10 cm length were interfaced with a pressure-flow system and a multimode fiber-loop cavity using concave fiber lenses with matching numerical apertures and diameters. Two red dyes, Allura Red AC and Congo Red, were detected with a 532 nm pulsed laser at a 5 nM limit of detection in a detection volume of less than 1 μL, corresponding to a minimal detectable absorbance of less than 4 × 10−4 cm−1 and a minimal detectable change in absorption cross section, σmin = Vdet × ε × CLOD, of about 14 μm2 (Allura Red AC) and 37 μm2 (Congo Red).