Abstract
Image-guided surgeries employ advanced imaging and computing technologies to assist the surgeon when direct visualization is inadequate or unavailable. As modern surgeries continue to move toward minimally invasive procedures, tracked ultrasound (US), an emerging technology that uniquely combines US imaging and position tracking, has been increasingly used for intraoperative guidance in surgical interventions.
The intrinsic accuracy of a tracked US system is primarily determined by a unique procedure called ``probe calibration", where a spatial registration between the coordinate systems of the transducer (provided by a tracking device affixed to the probe) and the US image plane must be established prior to imaging. Inaccurate system calibration causes misalignments between the US image and the surgical end-effectors, which may directly contribute to treatment failure. The probe calibration quality is further reduced by the "elevation beamwidth" or "slice thickness", a unique feature of the ultrasound beam pattern that gives rise to localization errors and imaging uncertainties.
In this thesis, we aim to provide an automated, pure-computation-based, intraoperative calibration solution that also incorporates the slice thickness to improve the calibration accuracy, precision and reliability. The following contributions have been made during the course of this research. First, we have designed and developed an automated, freehand US calibration system with instant feedback on its calibration accuracy. The system was able to consistently achieve submillimeter accuracy with real-time performance.
Furthermore, we have developed a novel beamwidth-weighted calibration framework (USB-FW) that incorporates US slice thickness to improve the estimation of calibration parameters. The new framework provides an effective means of quality control for calibration results. Extensive phantom validation demonstrated that USB-FW introduces statistically significant reduction (p = 0.001) in the calibration errors and produces calibration outcomes that are less variable than a conventional, non-beamwidth-weighted calibration.
Finally, we were the first to introduce an automated, intraoperative Transrectal Ultrasound (TRUS) calibration technology for needle guidance in prostate brachytherapy. Our tests with multiple commercial TRUS scanners and brachytherapy stepper systems demonstrated that the proposed method is practical in use and can achieve high calibration accuracy, precision and robustness.
URI for this record
http://hdl.handle.net/1974/7614Request an alternative format
If you require this document in an alternate, accessible format, please contact the Queen's Adaptive Technology CentreRelated items
Showing items related by title, author, creator and subject.
-
Enabling Indoor Location-based Services Using Ultrasound
Javed, Tayyab (2013-02-01)In the context of location, large amounts of information are available on the Internet to be accessed by people via different devices. However, at times people have to manually search and access it. If the space where ... -
Stochastic Modeling of Temporal Ultrasound Data for Prostate Cancer Diagnostics
Nahlawi, LayanDespite recent advances in clinical oncology, prostate cancer remains a major health concern in men, where current detection techniques still lead to both over- and underdiagnosis. More accurate prediction and detection ... -
Single Nucleotide Polymorphism of B-Type Natriuretic Peptide and Carotid Ultrasound Imaging as Biomarkers to Determine the Severity of Atherosclerosis
Li, TerryCardiovascular disease (CVD) has been the number one cause of death in the world for several decades and will continue to worsen as a global epidemic. The underlying cause of CVD is atherosclerosis: most myocardial infarctions ...